K Na Cl TCO2 Ion Selective Electrode ISE Analyzer with Color Touch Screen - Product
Place of Origin : Shenzhen,China
Delivery Time : 7 Working Days After Received Your Payment
Product Description
Ion Selective Electrode ISE Analyzer with Color Touch Screen
Main Features:
7" Color touch screen, easy to operate
High accurate and long life electrode and TCO2 sensor
Programmable multi-format print-out
Reagent pack, real time monitoring of reagent residual volume
Up to 50,000 test results can be stored
RS-232 port, supporting bar code reader and USB interface
Sleep mode to reduce reagent consumption
Option: Auto Sampler
Parameter configuration:
A K,
Physical Specification
Other Specification :
Ion Selective Electrode ISE Analyzer With Color Touch Screen
Main Features:
7" Color Touch Screen, Easy To Operate
High Accurate And Long Life Electrode And TCO2 Sensor
Programmable Multi-format Print-out
Reagent Pack, Real Time Monitoring Of Reagent Residual Volume
Up To 50,000 Test Results Can Be Stored
RS-232 Port, Supporting Bar Code Reader And USB Interface
Sleep Mode To Reduce Reagent Consumption
Option: Auto Sampler
Parameter Configuration:
A K, Na, Cl
B K, Na, Cl, TCO2
C K, Na, Cl, ICa, NCa, TCa, PH
D K, Na, Cl, ICa, NCa, TCa, PH, TCO2, AG
F K, Na, Cl, Li
G K, Na, Cl, Li, TCO2
H K, Na, Cl, ICa, NCa, TCa, PH, Li
I K, Na, Cl, ICa, NCa, TCa, PH, Li, TCO2, AG
J K, Na, Cl, ICa, NCa, TCa, PH, Mg, TCO2, AG
K K, Na, Cl, ICa, NCa, TCa, PH, Li, Mg, TCO2, AG
Specifications:
Sample: Serum, Plasma, Whole Blood, Cerebrospinal Fluid And Dilute Urine
Measuring Speed: 25s
Analysis Method: Ion Selective Electrode (ISE)
Sample Volume: 60-300ul
Sampler Position: 39 Positions (including 5 Emergency And 2 QC)
Sample Tube: Primary Tube, Micro Tube, Plastic Tube, Etc.
Storage: Up To 10,000 Test Results
Printer: Internal Thermal Printer
Interface: RS232 Port
Working Environment:
Temperature: 5-40
Relative Humidity: 80 %
Atmospheric Pressure: (86~106) KPa
Power Supply: 220V20V, 50-60Hz
Power: 120W
Dimension: 340mm*200mm*380mm
Net Weight: 6Kg
- - - -
Items | Measuring Range | Resolution | Measuring Precision (CV%) |
K+ | 0.5-20.0mmol/L | 0.01 Mmol/L | 1.0% |
Na+ | 15-200mmol/L | 0.1 Mmol/L | 1.0% |
Cl- | 15-200mmol/L | 0.1 Mmol/L | 1.0% |
Ca2+ | 0.1-6.0mmol/L | 0.01 Mmol/L | 1.0% |
Li+ | 0.1-5.0mmol/L | 0.01 Mmol/L | 2.0% |
PH | 4-9pH | 0.01 PH | 0.5% |
TCO2 |