multi-parameters patient monitor - Product
Place of Origin : Beijing,China
Payment Terms : T/T
Product Description
he parameters that the JB 9107 covers include: ECG, IBP(2 Channels),
NIBP, SPO2, Respiration, Pulse Rate, Temperature and
ETCO2. JB-9107 employs a 12” display to give detailed and clear
view to the caregiver. All the menus can be promptly accessed using
rotating knob, ensuring functions to be delivered decently during
operation. The built-in battery allows the monitor operated in patient
transfer and in case of an outage. Sophisticated networking design
gives the 9107 the abilit
Basic Information
Brand : JB
Physical Specification
Weight : 5kgs
Other Specification : Type :Portable
Application :Adult, Pediatric And Neonatal Patients
Classification :Types Of Protection: Class I Equipment
Degree Of Protection: Type CF Defib-proof
Safety: EN 60601-1
Display :Screen: 12.1” Full Color TFT
Resolution: 800*600
Numerics: HR, ST Segment, PR, NIBP, (Systolic,Diastolic, MAP And PR), IBP1, IBP2, BR,
EtCO2, Temp, %SpO2,
Waveforms: Up To 6 Waveforms: ECG, IBP1,IBP2, Resp, CO2, SPO2
Sweep Speed: 12.5, 25, 50 Mm/sec (ECG, SpO2) 6.25 Mm/sec (Resp, CO2)
Trend: Maximum 72 Hours
Controls
Function Keys: 6; Soft-moving Buttons
Rotary Knob: Push And Rotate; 24 Steps/turn
Alarm
Audible : Volume Adjustable
Visual: Figures Flashing
Networking
Ethernet LAN To 9100 Central Monitoring System
Recorder
Built-in, Thermal Array
Width: Paper/Chart 50mm/48mm
Paper Speed: 25mm/s
ECG
Five-lead And Twelve-lead Selectable: GND; Limb Leads I, II, III; Augmented Limb Leads AVL,
AVR, AVF; Precordial Leads V1, V2, V3, V4,
V5,V6
Displayed Waveforms: 2 Channels
Sweep Speed: 12.5mm/s, 25mm/s, 50mm/s
Gain Selection: 6 Levels, Plus Auto
HR Range
Adult 20 300bpm
Pediatric 20 350bpm
Neonatal 20 350bpm
Working Modes: Monitoring/Diagnostic/Surgery
Test Signal: 1mV(displayed Via Touchkey)
ST Detection: Yes
Measuring Range: -2.0 ~2.0mv
Alarm Preset: -2.0 ~ 2.0mv
Overrun Alarm: Audible And Visual Alarm Will Function When A Preset Value Is Met.
Electrosurgery Protection: Yes
Defibrillator Protection: Yes
CMRR:
Monitoring Mode: >105dB
Diagnostic Mode: >105dB
Operating Mode: > 105dB
Amplifier Bandwidth: 0.05 ~ 100Hz
NIBP
Technique: Oscillometric
Measuring Parameters: Systolic, Diastolic, MAP, Pulse Rate.
Operation Modes: Manual/Automatic/STAT
Automatic Measurement Cycles: 1 ~ 90 Minutes; Factory Set: 5 Minutes
Unit: MmHg/kPa Selectable
Automatic Cuff Deflation:
If Cuff Pres