Full Automatic Biochemistry Analyzer (YJ-180) - Service
Place of Origin : Zheng Zhou,China
Product Capacity : Biochemistry Analyzer
Payment Terms : LLC/TT/PAYPAL
Product Description
Model YJ-180
Analyze Unit Throughput Constant 180T/h (non-ISE)
Theory Absorption spectrometry, transmission turbidimetry
Methodology End-point \Fixed-time\Kinetic\Ion selective electrode
Test according set up, random access Reaction plate Contain 81 reaction cuvettes
Maximum reaction time 15 minutes
Minimum reaction volume 150 ul
Material of cuvettes Special UV-transparent optical glass
Cuvettes use Automatically cuvettes washing
Optical Unit Light Source Halogen-tungsten lamp
Basic Information
Brand : Foremed
Model : YJ-180
Color : WHITE
Physical Specification
Size or Volume : 150 UL
Other Specification : Model YJ-180
Analyze Unit Throughput Constant 180T/h (non-ISE)
Theory Absorption Spectrometry, Transmission Turbidimetry
Methodology End-point \Fixed-time\Kinetic\Ion Selective Electrode
Test According Set Up, Random Access Reaction Plate Contain 81 Reaction Cuvettes
Maximum Reaction Time 15 Minutes
Minimum Reaction Volume 150 Ul
Material Of Cuvettes Special UV-transparent Optical Glass
Cuvettes Use Automatically Cuvettes Washing
Optical Unit Light Source Halogen-tungsten Lamp
Photometer Reversed Optics
Wavelength 340nm-900nm Wavelength Option
Absorption 0-4 Abs
Reagent Unit Reagent Plate 60 Reagent Positions ,support For Single And Double Reagent, Reagent Completely Open
Reagent Storage Temperature 2~10 Degree
Reagent Bottle 15ml, 20ml, 30ml, 50ml Or 70ml ,dead Volume Is Less Than 1.0 Ml
Reagent Volume 20-500 Ul, Step By 1.0 Ul
Reagent/Sample Probe Automatically Checked ;Patent Technology Liquid Level Detection;Collision Protection Probe Cleaning:Automatic Washing For Both Inside And Outside.
Sample Unit Sample Plate 60 Sample Positions
Sample Type Human (animal) Serum \plasma\urine Etc.
Samples Cup Support One-off Micro Cups
Sample Volume 2-100 Ul, Step By 0.1 Ul
Sample/reagent Probe Automatically Checked ;Patent Technology Liquid Level Detection;Collision Protection Probe Cleaning:Automatic Washing For Both Inside And Outside.
Temperature Control Unit Heating Method Direct Thermal Solid Thermostat Closed
Temperature Equilibrium Time <3 Minutes
Reaction Temperature 37 ±0.1 Degree
Temperature Fluctuation ±0.1 Degree
Operation Unit Software WindowsXP \2000
Database SQL Server
Data Storage Automatic Store & Copy; No Time And Volume Limit; Lifetime Preservation
Print Format Can Be Set Up, Offer Several Standard Format
Calibration Mode Linear Curve Type: Single-point Linear, Two-point Linear And Multi-point Linear None Linear; Curve Type: Logistic-Log 4P,Logistic-Log5p